Arterial Line - Square Wave Test 
The arterial line can measure BP inaccurately unless properly calibrated. Rapidly flushing the line (by pulling the release on the flush device) generates a square wave. Counting oscillations after the square wave indicates if the arterial line is working properly.
Normal = accurate BP
    - 1.5 - 2 oscillations
OVERdamped = falsely LOW BP
    - <1.5 oscillations
    - Absence of dicrotic notch
    - Fix: Remove extra tubing & any air bubbles.
UNDERdamped = falsely HIGH BP
    - >2 oscillations
    - Additional artifactual spikes
    - Fix: Adjust filter settings on monitor

Nick Mark MD @nickmmark

#squarewave #test #dampening #damped #arterial #artline #diagnosis 